QTOC 202 Spezification
Analyser
| Mode of operation |
- thermal-catalytic oxidation of all carbon compounds at 800°C
- slight acidification of samples automatically
- exhalation of the inorganic C-compounds automatically
- measuring of CO2-contents by NDIR-detector (type ANCO 3)
- *measuring of TNb-contents (optionally)
- measuring of the total nitrogen by EC-sensor or CLD-detector
- low temperature reactor for measuring of the inorganic carbon at 160 °C
- possibility of changeover for analysis of solids
|
| Gas-analyser |
two comfortable NDIR-laboratory analysers with 3 reversible measuring ranges each guarantee:
- noise free measuring signals
- no double peaks
- high reproducibility
6 measuring ranges, reversible
- 0 – 50 / 0 – 200 / 0 – 1000 ppm CO2
- 0 – 1% / 0 – 5% / 0 – 10% CO2
other measuring ranges on request
|
| Analysing time |
from 30 sec. on per measuring |
| Autom. sampler (KSP) |
for fully automatic sample input sample rack for 40 (15 ml) resp. 104 (4 ml) single samples |
| Sample quantity |
20 - 400 µl |
| Particle transmissibility |
up to 500 µm |
| Temperature stabilisation |
A cooling device takes care of always non-varying temperatures for the 40 or 104 samples |
| Injection system |
Innovative injection system which prevents seeping of water samples |
| Reactor temperature |
800°C for TOC, 160°C for TIC, 2 identical furnaces, interchangeable |
| System-measuring range |
0 – 1000 mg/l C und 0 –200 mg/l N |
| Detection limit |
0,1 mg/l C (and 0,1 mg/l N with TNb -option) |
| Quality assurance |
accord. to EN 1484 / DIN 38409 / DIN 32645 / accord. to QTOC-data file 10.2004 |
| Analysis |
Graphics and log at external computer-analysing system |
| Main supply |
230 V / 50 Hz / 6 A |
| QTOC 201 dimensions |
725 x 520 x 600 mm (H x W x D) |
| Weight |
about 48 kg |
| Carrier gas supply |
Synthetic air or O2, class 4, 8 or better. Flow adjustable 4 – 20 l/h (manually) |
Solids Module
| Mode of operation |
- thermal-catalytic oxidation of the carbon-compounds at 1350 °C
- thermal-catalytic postoxidation at 800 °C
- exhalation of the inorganic carbon-compounds
- measuring of the CO2-contents by NDIR-detectors in the analyser type ANCO3 for solid and liquid phase
|
| Sample input |
closed sample-input system, automatically |
| Sample portion |
max. 5 g in ceramic boats in the solids section |
| Thermal fusion unit |
2-level-laboratory furnace. First heating zone: 1350 °C, second heating zone: 800 °C (for postoxidation), type LABOF |
| Temperature stabilisation |
highly exact automatic temperature control |
Quality assurance |
acc. to EN 1484 / DIN 38409 / DIN 32645 / EN 13137 |
| Main supply |
230 V / 50 Hz / 6 A |
| Dimensions |
300 x 370 x 660 mm (H x W x D) |
| Weight |
about 22 kg |
| Carrier gas supply |
supply by analysing unit |
| Measuring ranges |
3 measuring ranges: 0 – 1% / 0 – 5% / 0 – 10% CO2 |
| *TIC-module |
for acidification and exhalation at TIC-measuring (samples with inorganic C-compounds), optionally |
|
|