LABOF Spezification
Solids Module
| Mode of operation |
- thermal-catalytic oxidation of the carbon-compounds at 1350 °C
- thermal-catalytic postoxidation at 800 °C
- exhalation of the inorganic carbon-compounds
- measuring of the CO2-contents by NDIR-detectors in the analyser type ANCO3 for solid and liquid phase
|
| Sample input |
closed sample-input system, automatically |
| Sample portion |
max. 5 g in ceramic boats in the solids section |
| Thermal fusion unit |
2-level-laboratory furnace. First heating zone: 1350 °C, second heating zone: 800 °C (for postoxidation), type LABOF |
| Temperature stabilisation |
highly exact automatic temperature control |
Quality assurance |
acc. to EN 1484 / DIN 38409 / DIN 32645 / EN 13137 |
| Main supply |
230 V / 50 Hz / 6 A |
| Dimensions |
300 x 370 x 660 mm (H x W x D) |
| Weight |
about 22 kg |
| Carrier gas supply |
supply by analysing unit |
| Measuring ranges |
3 measuring ranges: 0 – 1% / 0 – 5% / 0 – 10% CO2 |
| *TIC-module |
for acidification and exhalation at TIC-measuring (samples with inorganic C-compounds), optionally |
|
|
© KURSCHAT GmbH, Lüneburger Schanze 16, D-21614 Buxtehude, Germany Tel.: +49 4161 / 7153-0, eMail: info@kurschat-gmbh.de

|